Differences in Blood Glucose Levels Between POCT (Point of Care Testing) and Spectrophotometer Methods
DOI:
https://doi.org/10.56359/igj.v4i3.645Keywords:
Blood Glucose, POCT (Point Of Care Testing), SpectrophotometerAbstract
Background & Objective: Blood glucose is a sugar found in the blood that is formed from carbohydrates in food and stored as glycogen in the liver and muscles. Glucose is one of the important carbohydrates used as the main source of energy in the body. Blood glucose test results exceeding the normal level of 200 mg/dl may indicate that a person has diabetes mellitus. This study aims to determine the difference in blood glucose levels between the POCT method and the spectrophotometer.
Method: The study design is analytical, using random sampling with 30 participants, and statistical analysis was performed using the Paired T-Test and Wilcoxon test.
Result: The results of this study using the Paired T-Test showed that POCT data is normally distributed (sig value = 0.208 > 0.05), while spectrophotometer data is not normally distributed (sig value = 0.003 < 0.05). Meanwhile, the Wilcoxon test yielded a sig value of 0.000 < 0.05.
Conclusion: The results of blood glucose level measurements using the POCT method showed an average value of 128.83 mg/dl, while the spectrophotometer results showed an average value of 88.07 mg/dl. Based on the statistical test results, the sig value was 0.000 (p<0.5), meaning H0 was rejected. This means that the difference between the two measurement methods is statistically significant.
Downloads
References
Anisa, Y. (2022). Analisis Perbedaan Hasil Pemeriksaan Glukosa Darah Menggunakan Gukometer Dan Spektrofotometer Pada Ibu Hamil Trimester Ii Dan Iii. Universitas Binawan.
Farmasi, J. S. (2018). Survei_Risiko_Penyakit_Diabetes_Melitus. 5(2), 134–141.
Fatin, Y. (2018). Pemeriksaan Glukosa Dengan Point Of Care Testing (Poct). Bukit Tinggi:Rumah Sakit Stroke Bukit Tinggi. Https://Repo.Upertis.Ac.Id/1535/1/Maini Roza.Pdf
Firgiansyah, A. (2016). Perbandingan Kadar Glukosa Darah Menggunakan Spektrofotometer Dan Glukometer. Fakultas Ilmu Keperawatan Dan Kesehatan Universitas Muhammadiyah Semarang, 13irgiansy(1), 1–71.
Gusmayani, Y., Anggraini, H., & Nuroini, F. (2021). Perbedaan Kadar Kolesterol Serum Metode Spektrofotometri Dan Metode Point Of Care Testing (Poct). Jurnal Labora Medika, 5(1), 24–28.
Mariady, F., Sugiarto, C., Sadeli, L., Mariady, F., Sugiarto, C., & Sadeli, L. (2013). Perbandingan Hasil Pemeriksaan Kadar Glukosa Darah Sewaktu Menggunakan Glukometer Dan Spektrofotometer Pada Penderita Diabetes Melitus Di Klinik Nirlaba Bandung Fakultas Kedokteran Universitas Kristen Maranatha Jl . Prof . Drg . Suria Sumantri No . 65 , B. Bandung, Fakultas Kedokteran Universitas Kristen Maranantha, 13(1), 1–8.
Maulidiyanti, E. T. S. (2017). Perbedaan Hasil Pemeriksaan Kadar Glukosa Darah 2 Jam Pp Dengan Menggunakan Glukometer Dan Analyzer Pada Penderita Diabetes Melitus. The Journal Of Muhammadiyah Medical Laboratory Technologist, 1(1), 16. Https://Doi.Org/10.30651/Jmlt.V1i1.978
Nining Kurniati, Putri Wulandari, & Wawan Sofwan Zaini. (2022). Perbandingan Hasil Pemeriksaan Gula Darah Sewaktu Metode Poct Dan Metode God-Pap Pada Pasien Diabetes Mellitus Dengan Nilai Kritis Di Igd Rsau Dr. M. Hassan Toto Bogor. Jurnal Teknologi Kesehatan Borneo, 3(2), 76–80. Https://Doi.Org/10.30602/Jtkb.V3i2.251